The optical chain of a POCT hemoglobin analyzer is surprisingly simple: light source -> interference filter -> 10 uL of whole blood -> photodetector -> readout conversion. No spectrometer, no detector array. The accuracy of the entire instrument ultimately rests on a few square millimeters of coated glass. And this make-or-break filter is a standard deliverable of the Bena Optics precision coating line: from thin-film design and coating processing to piece-by-piece inspection, all in one stop.
The reason lies in the Beer-Lambert law: absorbance A = ε(λ)·c·l, where the molar extinction coefficient ε is a function of wavelength. Once the passband widens, ε is no longer constant across it, and A loses its strict linear relationship with concentration c. The effect is barely visible at low concentrations, but errors grow systematically on high-value samples. For hemoglobin measurement, therefore, the filter's mission is not "the more light, the better" - it is to let only the right light through, and block everything that should not pass. That is a rule honed over more than a decade of Bena Optics coating experience: every layer thickness in the thin-film design serves one goal - a clean spectral window.

The Q-band features of hemoglobin are well defined: oxyhemoglobin (HbO₂) shows a double peak at about 542 nm and 576 nm, with an absorption valley at 560 nm between them; deoxyhemoglobin (Hb) shows a single peak at about 556 nm; the product of the HiCN method has its maximum absorption at 540 nm; MetHb and COHb have their characteristic region near 630 nm; and 805 nm is the isosbestic point of Hb and HbO₂. From these, engineers define the key wavelengths: 540 nm (ICSH reference method), 570 nm (main measurement for the azide-methemoglobin method), 880 nm (turbidity compensation), 630 nm (derivative discrimination), and 805 nm (reference).
This batch of interference filters for hemoglobin analyzers (QZ-1 through QZ-5) was coated by Bena Optics using a Ta₂O₅/SiO₂ hard-coating system on 2 mm fused silica substrates. After fabrication on high-precision coating machines, each piece underwent a full 400-1100 nm spectral scan. Laid out across the full spectrum, the passbands of the five filters line up almost one-to-one with the key wavelengths above:
Sample | Measured Passband | Peak | Role |
QZ-3 Green | 518-588 nm (center 561) | 95.13% | Main measurement: covers all Q-band features at 542/555/576 nm, including the 570 nm measuring wavelength |
QZ-5 Long-pass | >767 nm | 95.70% | Turbidity compensation: one channel covering the 805 nm isosbestic point and the 880 nm compensation wavelength |
QZ-4 Red | 631-691 nm (center 637) | 95.02% | Derivative channel: aligned with the MetHb/COHb region at 630 nm |
QZ-2 Blue | 450-516 nm | 95.02% | Blue-region auxiliary: covers the HiCN valley at 504 nm |
QZ-1 Broadband | 450-794 nm | 95.69% | White-light reference: covers the entire visible region |
When a filter arrives, most people first glance at the peak transmittance - yet it is precisely the least important metric. A slightly lower transmittance only costs you optical flux, which can be recovered by raising the source current or extending the integration time. The next two metrics are systematic errors that no back-end circuit can fix.
Out-of-band rejection determines the stray-light floor and is the most underestimated of the three metrics. QZ-2 averages only 0.0068% transmittance across its 530-780 nm stopband, and QZ-3 only 0.0090% across 600-1000 nm - both effectively at the instrument noise floor, i.e., "pitch black." Such a low leakage level comes from Bena Optics' exacting control of layer thickness: high-precision coating machines with real-time monitoring press every filter's stopband to the same level.
Cutoff steepness determines channel isolation. QZ-2 drops from 94% to 1.37% in only about 5 nm, a steepness of 20.1 %/nm; QZ-4 also achieves 19.8 %/nm on its short-wavelength side. Such steep edges come from the refractive index contrast of nearly 0.6 between Ta₂O₅ (n ≈ 2.1) and SiO₂ (n ≈ 1.46): with the right material pair and enough layers, high reflectance can be built with fewer stacks. Bena Optics optimizes the material combination and layer count through full-spectrum simulation at the design stage, ensuring every delivered piece maintains the same steepness.
Passband flatness determines the linear range. The main-measurement QZ-3 has an in-band standard deviation of 1.70, a reasonable level; QZ-5 shows the largest range of 15.75, but that is acceptable because it serves as the white-light reference. For a high-volume coating line, keeping in-band ripple stably within this window tests the consistency of coating equipment and the discipline of fixed process parameters.
QZ-3 averages only 0.0090% transmittance across its 600-1000 nm stopband, yet it surges after 1040 nm, reaching 71.24% at 1085 nm (with an additional 59.19% side peak at 1059-1067 nm). In other words, its stopband "ends early" at 1000 nm. Such side peaks are almost impossible to catch with the routine "spot-check of a few characteristic wavelengths" - they only surface when the full 400-1100 nm spectrum is laid out completely.
A silicon photodiode responds all the way out to 1100 nm, so this side peak falls squarely within the detection range. With a broadband source such as a tungsten-halogen lamp, this light passes through the green channel unimpeded and inflates the signal - presenting as lower absorbance and falsely elevated low-end readings that drift as the source ages, and that a single calibration cannot remove. There are three remedies: add a short-pass cut-off filter in front of the detector (the lowest cost), add a near-infrared absorbing filter at the source, or modify the thin-film stack to suppress the side peak beyond 1050 nm - the last option eliminates the problem at its source and is the permanent fix Bena Optics prefers.
A datasheet typically states only a single line - "stopband OD4 @ 200-1200 nm" - but exactly which wavelength leaks and by how much can only be revealed by a full-spectrum scan. That is why Bena Optics insists on a 400-1100 nm full-spectrum scan for every single filter rather than releasing products after spot-checking a few wavelengths. Surface inspection tools such as laser interferometers complement the process, ensuring that what we deliver is not just a "design value" but a verified "measured value."